Skip to main navigation Skip to search Skip to main content

A chip-based potentiometric sensor for a Zika virus diagnostic using 3D surface molecular imprinting

  • Stony Brook University
  • Codagenix Inc.
  • Polytechnic University

Research output: Contribution to journal › Article › peer-review

31 Scopus citations

Abstract

The latest Zika virus (ZIKV) pandemic caused great international concern from explosively proliferating throughout the Americas. Currently, there is no vaccine to prevent Zika virus infection and available tests rely on antibodies or RNA. Unfortunately, antibody-based detection systems can result in false positive results and RNA-based detection systems are costly, time-consuming, and impractical for testing in remote regions. In this study, a potential point-of-care (POC) diagnostic system was developed using a chip-based potentiometric sensor to detect Zika virus using a 3D molecular imprinting technique. This chip-based potentiometric sensor system was able to detect 10-1 PFU mL-1 ZIKV in a buffered solution under 20 minutes without any sample manipulation. This sensor was tested against Dengue virus at clinical viral loads and showed no sign of cross-reactivity. When tested against human saliva samples containing clinical viral loads, this sensor was able to detect 10 PFU mL-1 ZIKV among the pool of bio-macromolecules. The high sensitivity and high selectivity demonstrated here proved that this lab-on-a-chip diagnostic has the potential to become a POC detection system for rapid and accurate screening of flaviviruses.

Original languageEnglish
Pages (from-to)4266-4280
Number of pages15
JournalAnalyst
Volume144
Issue number14
DOIs
StatePublished - Jul 21 2019

Fingerprint

Dive into the research topics of 'A chip-based potentiometric sensor for a Zika virus diagnostic using 3D surface molecular imprinting'. Together they form a unique fingerprint.

Cite this