Abstract
Investigations of interpolating cathode “pad” readout have been carried out for high multiplicity, two dimensional position sensitive detection of minimum ionizing particles and heavy ions. A detector configuration representing only 0.6% of a radiation length and covering an area of 26 cm by 16 cm, with over 1000 readout channels, has been developed. In a prototype chamber using this technique, a resolution of less than 100 µm (rms), for 5.4 keV x-ray, and differential non-linearity of ±6% have been achieved. The technique has also been demonstrated for detectors of much larger area.
| Original language | English |
|---|---|
| Pages (from-to) | 88-94 |
| Number of pages | 7 |
| Journal | IEEE Transactions on Nuclear Science |
| Volume | 37 |
| Issue number | 2 |
| DOIs | |
| State | Published - Apr 1990 |
Fingerprint
Dive into the research topics of 'A study of wire chambers with highly segmented cathode pad readout for high multiplicity charged particle detection'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver