Abstract

We describe a MEMS-on-CMOS microsystem to encage, culture, and monitor cells. The system was designed to perform long-term measurements on arrays of single electrically active cells. A MEMS process flow was developed for the fabrication of closeable micro-vials to contain each cell, a custom bio-amplifier CMOS chip was designed, fabricated, and tested, and the fabrication of the MEMS structures on this chip was demonstrated. In addition, bovine aortic smooth muscle cells were plated on the surface, and over the course of a week they adhered, formed processes, and reproduced, verifying the compatibility of the materials used with the cell culture.

Original languageEnglish
Pages (from-to)2534-2537
Number of pages4
JournalAnnual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings
Volume26 IV
StatePublished - 2004
EventConference Proceedings - 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2004 - San Francisco, CA, United States
Duration: Sep 1 2004Sep 5 2004

Keywords

  • Cells
  • CMOS
  • Lab-on-a-chip
  • MEMS
  • Polypyrrole

Fingerprint

Dive into the research topics of 'Cell-lab on a chip: A CMOS-based microsystem for culturing and monitoring cells'. Together they form a unique fingerprint.