Abstract
We compare analytically techniques for multiple-photon coincidence imaging in terms of spatial resolution, detector efficiency (or simply efficiency), and system sensitivity for a spherical water phantom (or simply sensitivity). One analysis consists of comparing photon-photon coincidence single-photon-emission computerized tomography (PP-SPECT) with angularly unconstrained, electronically collimated triple-photon coincidence imaging technique (TPCIT). A second analysis compares positron-emission tomography using time-of-flight information (TOF-PET) with angularly constrained, electronically collimated TPCIT. The angularly unconstrained TPCIT has similar spatial resolution, higher efficiency, and higher sensitivity, as compared with PP-SPECT. The angularly constrained TPCIT has similar spatial resolution, lower efficiency, and lower sensitivity, as compared with TOF-PET. However, the angularly constrained TPCIT can, for brain imaging, reduce the localization range of photon-emission sites along projection rays from about 20 cm to about 1 cm during data acquisition while TOF-PET currently reduces the range to about 7 cm.
| Original language | English |
|---|---|
| Pages (from-to) | 1282-1292 |
| Number of pages | 11 |
| Journal | IEEE Transactions on Nuclear Science |
| Volume | 37 |
| Issue number | 3 |
| DOIs | |
| State | Published - Jun 1990 |