Abstract

One of the most promising high resolution positron emission tomography detector designs comprises an array of small, optically isolated scintillator crystals each coupled to an independent photosensor, such as an avalanche photodiode (APD). However, crosstalk between crystals (due to Compton scatter, photoelectron escape, or incomplete optical isolation) can significantly degrade performance and is expected to increase as crystals become narrower to improve spatial resolution. Various measures of crosstalk have been determined for different configurations of 4x8 blocks of 2 mm x 2 mm lutetium oxyorthosilicate (LSO) crystals coupled to matched Hamamatsu APD arrays. Results indicate that ignoring crosstalk signals could lead to a ~ 30% loss in coincidence sensitivity for a tomograph using these detectors and a decrease in energy resolution of ~ 2 percentage points. Spatial and time resolution would not be significantly affected with most practical front-end architectures.

Original languageEnglish
Pages (from-to)362-366
Number of pages5
JournalIEEE Transactions on Nuclear Science
Volume50
Issue number3
DOIs
StatePublished - Jun 2003

Keywords

  • Biomedical nuclear imaging
  • gamma-ray detectors
  • optical crosstalk
  • positron emission tomography (PET)
  • scintillation detectors