Skip to main navigation Skip to search Skip to main content

Enhancing the Linear Dynamic Range in Multi-Channel Single Photon Detector beyond 7OD

  • Advanced BioMedical Machines Inc.
  • Stony Brook University

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

We present design, implementation, and characterization of a single photon detector based on 32-channel photomultiplier tube (PMT) sensor (model H7260-20, Hamamatsu). The developed high-speed electronics enables the photon counting with linear dynamic range (LDR) up to 108 count/s per detector's channel. The experimental characterization and Monte Carlo simulations showed that in the single-photon counting mode the LDR of the PMT sensor is limited by photon pulsewidth (current pulse) of 900 ps and a substantial decrease of amplitudes of current pulses for count rates exceeding 10-8 count/s. The multi-channel architecture of the detector and the developed firm/software allow further expansion of the dynamic range of the device by 32-fold using appropriate beam shaping. The developed single-photon counting detector was tested for the detection of fluorescence labeled microbeads in capillary flow.

Original languageEnglish
Article number7217794
Pages (from-to)7081-7086
Number of pages6
JournalIEEE Sensors Journal
Volume15
Issue number12
DOIs
StatePublished - Dec 1 2015

Keywords

  • fluorescence detection.
  • photomultipliers
  • single photon counting

Fingerprint

Dive into the research topics of 'Enhancing the Linear Dynamic Range in Multi-Channel Single Photon Detector beyond 7OD'. Together they form a unique fingerprint.

Cite this