Skip to main navigation Skip to search Skip to main content

First results from the BNL/Penn PET-MRI system for whole body rodent imaging at 9.4T

  • M. Budassi
  • , M. L. Purschke
  • , J. Fried
  • , T. Cao
  • , S. Stoll
  • , E. Gualtieri
  • , J. S. Karp
  • , P. O'Connor
  • , D. J. Schlyer
  • , C. L. Woody
  • , P. Vaska
  • Stony Brook University
  • Brookhaven National Laboratory
  • University of Pennsylvania

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

4 Scopus citations

Abstract

A new PET system has been developed for the purposes of simultaneous PET/MR whole body rodent imaging in conjunction with a Varian large-bore 9.4T MRI system with a commercial Insight birdcage coil. The detector and readout technology is based on that developed for the RatCAP PET system, resulting in a highly robust and compact design that is compatible with MRI systems at the highest field strength. Testing of the full readout chain has shown a successful implementation of both modalities, with indications of only modest cross-interference on MRI image quality or PET detection efficiency. The final design cycle is complete, and supports operation of the full system with upgrades to the data acquisition electronics, firmware, and software. First results from the complete PET system will be presented, including preliminary imaging and tests of interference between modalities.

Original languageEnglish
Title of host publication2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012
Pages2753-2755
Number of pages3
DOIs
StatePublished - 2012
Event2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012 - Anaheim, CA, United States
Duration: Oct 29 2012Nov 3 2012

Publication series

NameIEEE Nuclear Science Symposium Conference Record
ISSN (Print)1095-7863

Conference

Conference2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, NSS/MIC 2012
Country/TerritoryUnited States
CityAnaheim, CA
Period10/29/1211/3/12

Fingerprint

Dive into the research topics of 'First results from the BNL/Penn PET-MRI system for whole body rodent imaging at 9.4T'. Together they form a unique fingerprint.

Cite this