Skip to main navigation Skip to search Skip to main content

Improvement of energy resolution in geiger-mode APD arrays using curve-fitting of signal decay

  • Irving N. Weinberg
  • , Pavel Y. Stepanov
  • , Arielle S. Weinberg
  • , Pamela Abshire
  • , Marc Dandin
  • Weinberg Medical Physics LLC
  • University of Maryland, College Park

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

A method is presented to improve the energy resolution of scintillators read out with Geiger-mode avalanche photodiode arrays. The method employs digital signal processing, in which individual decay curves for gamma-ray detection events are digitized and then fitted to analytical functions whose amplitude provides energy information. Simulation studies suggest that after-pulses represent the largest source of energy resolution loss, which can be improved with curve-fitting. An experimental measurement confirmed that energy resolution could be improved with least-square curve-fits to a simple exponential model.

Original languageEnglish
Title of host publication2008 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2008
Pages1416-1418
Number of pages3
DOIs
StatePublished - 2008
Event2008 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2008 - Dresden, Germany
Duration: Oct 19 2008Oct 25 2008

Publication series

NameIEEE Nuclear Science Symposium Conference Record
ISSN (Print)1095-7863

Conference

Conference2008 IEEE Nuclear Science Symposium Conference Record, NSS/MIC 2008
Country/TerritoryGermany
CityDresden
Period10/19/0810/25/08

Fingerprint

Dive into the research topics of 'Improvement of energy resolution in geiger-mode APD arrays using curve-fitting of signal decay'. Together they form a unique fingerprint.

Cite this