Internal Cardioverter Defibrillator: Component Failure

  • Stony Brook University

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Sudden death occurred in a patient with congestive cardiomyopathy, sustained ventricular tachycardia refractory to amiodarone, and an automatic internal Cardioverter defibrillator. Interrogation of the AICD at the time of death was indicative of malfunction. Postmortem analysis documented component failure involving a voltage regulating transistor resulting in prolonged charge times.

Original languageEnglish
Pages (from-to)1086-1088
Number of pages3
JournalPACE - Pacing and Clinical Electrophysiology
Volume13
Issue number9
DOIs
StatePublished - Sep 1990

Keywords

  • automatic internal Cardioverter defibrillator
  • component failure
  • ventricular fibrillation