TY - GEN
T1 - Low-power EEG monitor based on compressed sensing with compressed domain noise rejection
AU - Bertoni, Nicola
AU - Senevirathna, Bathiya
AU - Pareschi, Fabio
AU - Mangia, Mauro
AU - Rovatti, Riccardo
AU - Abshire, Pamela
AU - Simon, Jonathan Z.
AU - Setti, Gianluca
N1 - Publisher Copyright:
© 2016 IEEE.
PY - 2016/7/29
Y1 - 2016/7/29
N2 - Wireless sensor nodes capable of acquiring and transmitting biosignals are increasingly important to address future needs in healthcare monitoring. One of the main issues in designing these systems is the unavoidable energy constraint due to the limited battery lifetime, which strictly limits the amount of data that may be transmitted. Compressed Sensing (CS) is an emerging technique for introducing low-power, real-time compression of the acquired signals before transmission. The recently developed rakeness approach is capable of further increasing CS performance. In this paper we apply the rakeness-CS technique to enhance compression capabilities for electroencephalographic (EEG) signals, and particularly for Evoked Potentials (EP), which are recordings of the neural activity evoked by the presentation of a stimulus. Simulation results demonstrate that EPs are correctly reconstructed using rakeness-CS with a compression factor of 16. Additionally, some interesting denoising capabilities are identified: the high-frequency noise components are rejected and the 60 Hz power line noise is decreased by more than 20 dB with respect to the state-of-the-art filtering when rakeness-CS techniques are applied to the EEG data stream.
AB - Wireless sensor nodes capable of acquiring and transmitting biosignals are increasingly important to address future needs in healthcare monitoring. One of the main issues in designing these systems is the unavoidable energy constraint due to the limited battery lifetime, which strictly limits the amount of data that may be transmitted. Compressed Sensing (CS) is an emerging technique for introducing low-power, real-time compression of the acquired signals before transmission. The recently developed rakeness approach is capable of further increasing CS performance. In this paper we apply the rakeness-CS technique to enhance compression capabilities for electroencephalographic (EEG) signals, and particularly for Evoked Potentials (EP), which are recordings of the neural activity evoked by the presentation of a stimulus. Simulation results demonstrate that EPs are correctly reconstructed using rakeness-CS with a compression factor of 16. Additionally, some interesting denoising capabilities are identified: the high-frequency noise components are rejected and the 60 Hz power line noise is decreased by more than 20 dB with respect to the state-of-the-art filtering when rakeness-CS techniques are applied to the EEG data stream.
UR - https://www.scopus.com/pages/publications/84983400316
U2 - 10.1109/ISCAS.2016.7527292
DO - 10.1109/ISCAS.2016.7527292
M3 - Conference contribution
AN - SCOPUS:84983400316
T3 - Proceedings - IEEE International Symposium on Circuits and Systems
SP - 522
EP - 525
BT - ISCAS 2016 - IEEE International Symposium on Circuits and Systems
PB - Institute of Electrical and Electronics Engineers Inc.
T2 - 2016 IEEE International Symposium on Circuits and Systems, ISCAS 2016
Y2 - 22 May 2016 through 25 May 2016