Abstract
Background: Metallosis is an uncommon phenomenon observed in late failures of cemented total knee arthroplasty (TKA), and it is rarely seen in the absence of metal-on-metal articulation. Methods: We report the case of a TKA patient with cemented titanium-alloy components and a polyethylene patella that was revised for early loosening with intra-operative severe metallosis. Results: We found that loosening and severe metallosis were associated with methacrylate particle abrasion on titanium alloy surfaces. Conclusion: Serum titanium ion level measurement may be helpful in the workup of a painful TKA with titanium-alloy components in order to establish a diagnosis.
| Original language | English |
|---|---|
| Pages (from-to) | 728-731 |
| Number of pages | 4 |
| Journal | Knee |
| Volume | 25 |
| Issue number | 4 |
| DOIs | |
| State | Published - Aug 2018 |
Keywords
- Metallosis
- Titanium
- Total knee arthroplasty
Fingerprint
Dive into the research topics of 'Metallosis in cemented titanium alloy total knee arthroplasty without apparent metal-on-metal articulation'. Together they form a unique fingerprint.Cite this