Misfit of three different implant-abutment connections before and after cyclic load application: An in vitro study

  • Sergio Alexandre Gehrke
  • , Rafael Arcesio Delgado-Ruiz
  • , Juan Carlos Prados Frutos
  • , María Prados-Privado
  • , Berenice Anina Dedavid
  • , Jose Manuel Granero Marín
  • , José Luiz Calvo Guirado
  • Biotecnos - Technology and Science
  • Pontifícia Universidade Católica do Rio Grande do Sul
  • Universidad Católica del Uruguay
  • Universidad Rey Juan Carlos
  • Universidad Católica San Antonio de Murcia

Research output: Contribution to journalArticlepeer-review

37 Scopus citations

Abstract

Purpose: This study aimed to evaluate the misfit of three different implant-abutment connections before and after cycling load. Materials and Methods: One hundred twenty dental implants and correspondent prefabricated titanium abutments were used. Three different implant-abutment connections were evaluated: Morse taper (MT group), external hexagon (EH group), and internal hexagon (IH group). Forty implants and 40 abutments were used per group. The parameters for the mechanical evaluation were set as: 360,000 cycles, load of 150 N, and frequency of 4 Hz. Samples were sectioned in their longitudinal and transversal axes, and the misfit of the implant-abutment connection was evaluated by scanning electron microscopy analysis. One-way analyses of variance, Tukey post hoc analyses (α = .05), and t test (P < .05) were used to determine differences between groups. Results: At the longitudinal direction, all the groups showed the presence of microgaps before cycling load; after cycling load, microgaps were reduced in all groups (P > .05). Transversally, only the MT group showed full fitting after cycling load compared with the other groups (EH and IH) (P < .0001). Conclusion: The application of cycling load produces an accommodation of the implant-abutment connection in internal, external, and Morse taper connections. In the longitudinal direction, the accommodation decreases and/or eliminates the gap observed initially (before load). In the horizontal direction, Morse cone implant-abutment connections experience a complete accommodation with the elimination of the gap.

Original languageEnglish
Pages (from-to)822-829
Number of pages8
JournalInternational Journal of Oral and Maxillofacial Implants
Volume32
Issue number4
DOIs
StatePublished - 2017

Keywords

  • Cyclic test
  • Dental implant
  • Implant-abutment interface
  • Implant-supported prostheses
  • Metallographic analysis