@inproceedings{724b801af64e4070b1bf207ba40b1816,
title = "Multi-Well Avalanche Selenium Detector for Time-of-Flight PET",
abstract = "For the first time, we propose using amorphous selenium (a-Se) as the photoconductive material for time-of-flight (TOF) PET detectors. One major drawback of a-Se is its poor timing jitter stemming from low carrier mobility due to shallow-traps, a problem that must be circumvented for TOF applications. Thus, we propose using a nanopattern multi-well selenium detector that enables both avalanche multiplication gain and ultrafast charge sensing in one device. We've previously shown the capabilities of our avalanche mode multi-well selenium detector, which has a large active area, impact ionization avalanche gain and timing jitter on the order of 100 ps. Using Cram{\'e}r-Rao Lower Bound analysis in conjunction with experimental measurements, we calculate the lower bound on coincidence time resolution for our selenium photodetector and compare it with that of the current state-of-the-art TOF-PET photodetectors. Our results show that our multi-well selenium detector is a promising candidate to achieve 100 ps coincidence time resolution in TOF-PET using standard-sized scintillator crystals with a photodetector that is cost-effective and uniformly scalable to large area.",
keywords = "Amorphous Selenium, Cram{\'e}r-Rao Lower Bound, CTR, PDE, PET, SiPM, SPTR",
author = "Andy Labella and Wei Zhao and Goldan, \{Amir H.\}",
note = "Publisher Copyright: {\textcopyright} 2019 IEEE.; 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2019 ; Conference date: 26-10-2019 Through 02-11-2019",
year = "2019",
month = oct,
doi = "10.1109/NSS/MIC42101.2019.9059675",
language = "English",
series = "2019 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2019",
publisher = "Institute of Electrical and Electronics Engineers Inc.",
booktitle = "2019 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2019",
}