Abstract
Precise manometric pressure, volume, and temperature (P-V-T) measurements of carbon in samples, standards, and blanks are critical for radiocarbon studies. While P and T uncertainties depend on instrument choice and environmental stability, V uncertainties depend on their method of measurement and are often overlooked. We used numerical simulations and error propagation to find optimum procedures for measuring cold-finger volumes equipped with capacitance diaphragm gauges (CDGs) by two common application of Boyle's Law: cryogenic transfers and serial gas expansions with a reference flask of known volume. Minimum relative uncertainties of cold-finger volumes are comparable for these two methods (∼0.002), but the serial gas expansion method is preferred due to its convenience. Serial gas expansions can be performed to high precision by using dry air, an initial pressure ∼76% full-scale (e.g., 760 Torr), and a reference flask with an optimal volume based on preliminary estimates of cold-finger volumes and an empirical power function. The volumes of cold-fingers ≥ 12 cm3 can be determined with minimum achievable relative uncertainties of 0.0021 to 0.0023. This limit translates to minimum achievable relative uncertainties of 0.0026 to 0.0027 for P-V-T measurements of moles of gas simulated here.
| Original language | English |
|---|---|
| Pages (from-to) | 1077-1089 |
| Number of pages | 13 |
| Journal | Radiocarbon |
| Volume | 61 |
| Issue number | 4 |
| DOIs | |
| State | Published - Aug 1 2019 |
Keywords
- capacitance diaphragm gauges
- manometry
- optimization
- uncertainty
- vacuum line
Fingerprint
Dive into the research topics of 'Optimized Volume Determinations and Uncertainties for Accurate and Precise Manometry'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver