Abstract

Objective: To report pressure and stiffness, in healthy volunteers, of a new compression device with an air bladder inflated by a pump to regulate pressure. Methods: The device was applied to 60 legs of 30 volunteers and set to exert different pressures of 20–50 mmHg. The exerted pressure was measured in supine and standing positions and during simple physical exercises; static stiffness index, dynamic stiffness index, and walking pressure amplitudes were calculated. Results: The exerted pressure showed a good correlation with the expected pressure at each pressure range. The stiffness indices were >10 mmHg in the range of inelastic materials. The device was considered very easy to apply and use by the testing researchers. Conclusions: The device stiffness is in the same range as the inelastic bandages. Consequently, similar hemodynamic effectiveness could be expected but must be proved. Unlike inelastic bandages, this device was easy to apply and use.

Original languageEnglish
Pages (from-to)456-464
Number of pages9
JournalPhlebology
Volume39
Issue number7
DOIs
StatePublished - Aug 2024

Keywords

  • Compression therapy
  • compression wraps
  • stiffness index

Fingerprint

Dive into the research topics of 'Pressure and stiffness of a new air-inflated compression wrap'. Together they form a unique fingerprint.