Skip to main navigation Skip to search Skip to main content

Proximal migration and compaction of an Enterprise stent into a coiled basilar apex aneurysm: a posterior circulation phenomenon?

  • Shervin R. Dashti
  • , David Fiorella
  • , Maria M. Toledo
  • , Yin Hu
  • , Cameron G. McDougall
  • , Felipe C. Albuquerque
  • St. Joseph's Hospital and Medical Center, Phoenix

Research output: Contribution to journalArticlepeer-review

25 Scopus citations

Abstract

A patient with a giant partially thrombosed basilar apex aneurysm was treated with balloon-assisted coil embolization. At the conclusion of the embolization, an Enterprise stent was placed from the upper basilar artery to the left P1 segment. Follow-up angiography 4 months later showed that the distal stent had pulled out of the posterior cerebral artery and was in the coil mass within the aneurysm. Angiography clearly showed that the stent had migrated down the basilar artery and that the artery was now kinked proximal to the stent. Nonetheless, the left posterior cerebral artery and the basilar artery were still widely patent. This is the first documented case in which a stent 'compacted' into a completed coiled aneurysm in a delayed fashion.

Original languageEnglish
Pages (from-to)356-358
Number of pages3
JournalJournal of NeuroInterventional Surgery
Volume2
Issue number4
DOIs
StatePublished - Dec 2010

Fingerprint

Dive into the research topics of 'Proximal migration and compaction of an Enterprise stent into a coiled basilar apex aneurysm: a posterior circulation phenomenon?'. Together they form a unique fingerprint.

Cite this