Skip to main navigation Skip to search Skip to main content

Spatial resolution improvement of PICOSEC Micromegas precise timing detectors

  • F. M. Brunbauer
  • , R. Aleksan
  • , Y. Angelis
  • , S. Aune
  • , J. Bortfeldt
  • , M. Brunoldi
  • , J. Datta
  • , D. Desforge
  • , G. Fanourakis
  • , D. Fiorina
  • , K. J. Floethner
  • , M. Gallinaro
  • , F. Garcia
  • , I. Giomataris
  • , K. Gnanvo
  • , Q. Huang
  • , F. J. Iguaz
  • , D. Janssens
  • , A. Kallitsopoulou
  • , I. Karakoulias
  • M. Kovacic, P. Legou, M. Lisowska, J. Liu, M. Lupberger, I. Maniatis, M. Micetic, H. Muller, E. Oliveri, T. Papaevangelou, M. Pomorski, L. Ropelewski, K. Salamon, D. Sampsonidis, L. Scharenberg, T. Schneider, E. Scorsone, L. Sohl, N. Shankman, M. van Stenis, Y. Tsipolitis, S. E. Tzamarias, A. Utrobicic, I. Vai, R. Veenhof, P. Vitulo, X. Wang, S. White, Z. Zhang, Y. Zhou
  • CERN
  • Université Paris-Saclay
  • Aristotle University of Thessaloniki
  • Ludwig Maximilian University of Munich
  • University of Pavia
  • National Institute for Nuclear Physics
  • Demokritos National Centre for Scientific Research
  • Laboratório de Instrumentação e Física Experimental de Partículas
  • University of Helsinki
  • Thomas Jefferson National Accelerator Facility
  • University of Science and Technology of China
  • University of Zagreb
  • University of Bonn
  • CEA Saclay
  • CIRI-AUTH
  • Stony Brook University
  • National Technical University of Athens
  • Ruder Boskovic Institute
  • University of Virginia

Research output: Contribution to journal › Article › peer-review

Abstract

The combination of a Cherenkov radiator with a semi transparent photocathode and a Micromegas based amplification stage allows PICOSEC Micromegas detectors to achieve a time resolution of better than 15 ps. While tileable prototypes with 10×10 channels feature 1 × 1 cm 2 readout pads, finer readout granularity can be used to improve the spatial resolution. We report on the study of high readout granularity PICOSEC Micromegas prototypes which achieve around 0.5 mm spatial resolution with 3.5 mm large pads. No significant improvement was found when readout pad size was further reduced to 2.2 mm. The timing resolution of the leading pad was found to be slightly degraded but remained better than 20 ps for a medium granularity prototype. The achieved spatial resolution can enable PICOSEC Micromegas to be used as precise timing and moderate resolution tracking detector simultaneously.

Original languageEnglish
JournalJournal of Instrumentation
Volume21
Issue number5
DOIs
StatePublished - May 2026

Keywords

  • Gaseous detectors
  • Micropattern gaseous detectors (MSGC, GEM, THGEM, RETHGEM, MHSP, MICROPIC, MICROMEGAS, InGrid, etc)
  • Timing detectors

Fingerprint

Dive into the research topics of 'Spatial resolution improvement of PICOSEC Micromegas precise timing detectors'. Together they form a unique fingerprint.

Cite this