Skip to main navigation Skip to search Skip to main content

Temperature Dependence of Retention in Reversed-Phase Liquid Chromatography. 2. Mobiie-Phase Considerations

  • Lynn A. Colet
  • , John G. Dorsey
  • , Ken A. Dill
  • , Ken A. Dill
  • University of Cincinnati
  • University of California at San Francisco

Research output: Contribution to journalArticlepeer-review

159 Scopus citations

Abstract

The retention mechanism In reversed-phase liquid chromatography (RPLC) has been examined over a wide temperature range with emphasis on the role of the mobile phase, van’t Hoff plot shapes were used to assess the retention mechanism, and the data showed evidence of the hydrophobic effect when water-rich and/or hydrogen-bonded mobile phases such as methanol/water were used. However, different van’t Hoff plot shape was observed with acetonltrlle/water mobile phases, Indicating a change In the retention mechanism. These data showed that the hydrophobic effect, which had previously been proposed as the driving force for retention, Is not a satisfactory explanation for the retention process in all RPLC systems.

Original languageEnglish
Pages (from-to)1324-1327
Number of pages4
JournalAnalytical Chemistry
Volume64
Issue number13
DOIs
StatePublished - Jul 1992

Fingerprint

Dive into the research topics of 'Temperature Dependence of Retention in Reversed-Phase Liquid Chromatography. 2. Mobiie-Phase Considerations'. Together they form a unique fingerprint.

Cite this