Abstract
We studied 33 clinically stable patients with permanent atrial fibrillation (AF), implanted with a ventricular demand rate-responsive (VVIR) pacemaker or an automatic defibrillator, in order to evaluate whether continuous right ventricular apex pacing (VP) conferring rate regulation may be advantageous when compared with slower drug-controlled AF. Devices were chronically programmed at ventricular backup pacing. Patients were divided in two groups according to their normal (n = 17) or depressed (n = 16) left ventricular systolic function (LVSF). Ventricular function was studied by using tissue Doppler and color M-mode and echocardiography, as well as B-type natriuretic peptide (BNP) measurements. Baseline data during AF were compared to corresponding measurements following a 1-month pacing period after the devices were programmed at a base rate of 70 beats/min. In both groups, VP worsened some indexes of left and right ventricular function (P < 0.05) without significantly affecting cardiac output, left ventricular filling pressures and BNP (P = not significant). We conclude that VP should not be considered advantageous compared to slower AF.
| Original language | English |
|---|---|
| Pages (from-to) | 403-408 |
| Number of pages | 6 |
| Journal | Heart and Vessels |
| Volume | 23 |
| Issue number | 6 |
| DOIs | |
| State | Published - 2008 |
Keywords
- Atrial fibrillation
- Ventricular function
- Ventricular pacing
Fingerprint
Dive into the research topics of 'Usefulness of rate regulation through continuous ventricular pacing in patients with drug-controlled slower atrial fibrillation and normal or depressed left ventricular systolic function'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver