Abstract
In order to model photon attenuation and detector resolution variation as a depth-dependent convolution for efficient reconstruction of quantitative SPECT, a central-ray approximation is necessary. This work investigates the impact of the approximation upon reconstruction accuracy and computational efficiency. A patient chest CT image was acquired and converted into an object-specific attenuation map. From a segmentation of the map, an emission thorax phantom was constructed with a cardiac insert. To generate a system-specific resolution-variant kernel, a point source was measured at several depths from the surface of a low-energy, high-resolution, parallel-hole collimator of a SPECT system. Projections of parallel-beam geometry were simulated from the phantom, the map, and the kernel on an elliptical orbit. Reconstruction was performed by the ML-EM algorithm with and without the central-ray approximation. The approximation cuts down dramatically (more than 100 fold) the computing time with a negligible loss (less than 1%) of reconstruction accuracy.
| Original language | English |
|---|---|
| Pages (from-to) | 433-439 |
| Number of pages | 7 |
| Journal | Physics in Medicine and Biology |
| Volume | 42 |
| Issue number | 2 |
| DOIs | |
| State | Published - 1997 |