Abstract
Neurotransmitter sensing is critical in studying nervous pathways and neurological disorders. A 16-channel current-measuring VLSI potentiostat with multiple ranges from picoamperes to microamperes is presented for electrochemical detection of electroactive neurotransmitters like dopamine, nitric oxide etc. The analog-to-digital converter design employs a current-mode, first-order single-bit delta-sigma modulator architecture with a two-stage, digitally reconflgurable oversampling ratio for ranging the conversion scale. An integrated prototype is fabricated in CMOS technology, and experimentally characterized. Real-time multi-channel acquisition of dopamine concentration in vitro is performed with a microfabricated sensor array.
| Original language | English |
|---|---|
| Pages (from-to) | 4063-4066 |
| Number of pages | 4 |
| Journal | Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings |
| Volume | 26 VI |
| State | Published - 2004 |
| Event | Conference Proceedings - 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2004 - San Francisco, CA, United States Duration: Sep 1 2004 → Sep 5 2004 |