Abstract
A new hard X-ray micro-spectroscopy beamline has recently been installed at bending-magnet beamline X27A at the National Synchrotron Light Source, where the focus of research is primarily directed towards the environmental, geological and materials science communities. This instrument delivers moderate, ∼10 μm spatial resolution using achromatic dynamically bent Kirkpatrick-Baez mirrors, in addition to providing high X-ray flux throughput and selectable energy resolution. The balance between moderate spatial resolution and high flux throughput, in combination with a liquid nitrogen-cooled 13-element energy-dispersive high-purity germanium detector, is particularly well suited to the investigation of dilute and thin-film systems using the fluorescence X-ray absorption fine-structure mode of detection. In this paper, we report on the design and performance of this instrument and highlight a recent experimental study undertaken at this facility.
| Original language | English |
|---|---|
| Pages (from-to) | 487-494 |
| Number of pages | 8 |
| Journal | Nuclear Inst. and Methods in Physics Research, A |
| Volume | 562 |
| Issue number | 1 |
| DOIs | |
| State | Published - Jun 15 2006 |
Fingerprint